• Order with calibration
  • Not sure what you need? Ask our Team

Additel 760-MA

Key Features

Enquire

A portable automated pressure calibrator in the palm of your hand—this could be our most exciting product yet! The Additel 760 series Automatic Handheld Pressure Calibrator takes portable pressure calibration to new levels. Weighing less than 4 lbs (1.8 kg), the ADT760’s innovative design contains a built-in pump, precision pressure sensor, internal controller and a large touch-screen color display. To generate pressure, simply key in the desired pressure and the Additel 760 will do the rest. Each unit has four channels: one internal pressure channel for source and measure pressures, two external pressure measurement channels, and one electronic measure and source channel. This series of calibrator has three standard models with the option of adding HART communications, documentation and data logging.

ADT760-MA
The 760-MA generates and controls pressure from vacuum pressures up to 300 psig (20 bar.g) with an accuracy of 0.02%FS. Equipped with a built-in barometric reference, each unit can switch between gauge and absolute pressure types. A variety of internal sensors are available which offer lower pressure ranges for improved performance.

Documenting Process Functionality
Each model of the Additel 760 series has an option incorporating documentation and communication functions turning your 760 into a multifunction documenting process calibrator. This feature provides HART communication, task documentation and data logging.

Pressure Specification
Max Pressure Range -12.5 to 300 psig (-0.86 to 20 bar.g)
Accuracy 0.02% FS
Stability <0.005% FS
Pressure Type Differential, Gauge
Over Range Indication 1.2
Resolution 6 digits
Measurement Units Pa, hPa, kPa, mPa, bar, mbar, msi, mmHg @ 0°C, cmHg @ 0°C, mHg @ 0°C, inHg @ 0°C, inH2O @ 4°C, mmH2O @ 4°C, cmH2O @ 4°C, mH2O @ 4°C, inH2O @ 20°C, mmH2O @ 20°C, cmH2O @ 20°C, mH2O @ 20°C, inH2O @ 68°C, kgf/cm2, mtorr, torr, lb/ft2, tsi, custom
Barometric Accuracy 60 Pa
Connection Hose, 5 ft (1.5 m), with built-in filter to 1/4BSPF, 1/4NPTF, and M20F adapters
Pressure Output Rate <90 Seconds (300 psi/5 ml)

 

Selection Guide - Differential Pressure
Module Module Range[6] Media Accuracy (%FS)[1] Burst Pressure 760-LLP 760-D 760-MA
inH2O mbar
DP025 ±0.25 ±0.62 G 0.2[2] 100x    
DP050 ±0.5 ±1.25 G 0.1[3] 100x    
DP1 ±1 ±2.5 G 0.05[4] 100x    
DP2 ±2 ±5 G 0.05[4] 100x    
DP5 ±5 ±10 G 0.05[4] 50x    
DP10 ±10 ±25 G 0.05[4] 20x  
DP20 ±20 ±50 G 0.05 20x  
DP30 ±30 ±75 G 0.05 20x  
DP50 ±50 ±125 G 0.05 3x    
DP100 ±100 ±250 G 0.02 3x    
DP150 ±150 ±350 G 0.02 3x    
DP300 ±300 ±700 G 0.02 3x    
DP400 -380 to 400 -950 to 1K G 0.02 3x    
DP800 -380 to 800 -950 to 2K G 0.02 3x    
DP1K -380 to 1K -950 to 2.5K G 0.02 3x    
[1] FS specification applies to the span of the module range. Accuracy includes one-year stability, except for DP025 to DP10 modules.
[2] Accuracy is a 6 months spec, 1-year long-term drift is 0.2%FS.
[3] Accuracy is a 6 months spec, 1-year long-term drift is 0.1%FS
[4] Accuracy is a 6 months spec, 1-year long-term drift is 0.05%FS.
[5] Specification based on gauge measurement. An additional 60 pa uncertainty will need to be included when measuring in absolute mode. Applicable only for use with the ADT760-MA
[6] The low module pressure range may be outside the pressure range of the calibrator
* ADT155 Pressure Modules are calibrated in psi and inH2O.

 

Selection Guide - Gauge Pressure
Module Module Range[6] Media Accuracy (%FS)[1] Burst Pressure 760-LLP 760-D 760-MA
psi bar
CP10 ±10 ±0.7 G 0.02[5] 3x  
CP15 -13.5 to 15 -0.95 to 1 G 0.02[5] 3x  
CP30 -13.5 to 30 -0.95 to 2 G 0.02[5] 3x  
CP35 -13.5 to 35 -0.95 to 2.5 G 0.02[5] 3x  
CP50 -13.5 to 50 -0.95 to 3.5 G 0.02[5] 3x    
CP100 -13.5 to 100 -0.95 to 7 G 0.02[5] 3x    
CP150 -13.5 to 150 -0.95 to 10 G 0.02[5] 3x    
CP200 -13.5 to 200 -0.95 to 14 G 0.02[5] 3x    
CP300 -13.5 to 300 -0.95 to 20 G 0.02[5] 3x    
[1] FS specification applies to the span of the module range. Accuracy includes one-year stability, except for DP025 to DP10 modules.
[2] Accuracy is a 6 months spec, 1-year long-term drift is 0.2%FS.
[3] Accuracy is a 6 months spec, 1-year long-term drift is 0.1%FS
[4] Accuracy is a 6 months spec, 1-year long-term drift is 0.05%FS.
[5] Specification based on gauge measurement. An additional 60 pa uncertainty will need to be included when measuring in absolute mode. Applicable only for use with the ADT760-MA
[6] The low module pressure range may be outside the pressure range of the calibrator
* ADT155 Pressure Modules are calibrated in psi and inH2O.

 

Electrical Specification
  Range Resolution Accuracy Note
mA Measure ±30 mA 0.0001 mA 0.01%RD+0.005%FS Impedance<10Ω
V Measure ±30 V 0.0001 V 0.01%RD+0.005%FS Impedance >1MΩ
mA Source 24 mA 0.001 mA 0.01%RD+0.005%FS 20 mA @ 1K
Loop Power Source 24 V N/A ±1 V 50 mA (Max Loading)
Pressure Switch Open, close. Support for mechanical switches and NPN/PNP digital switches.
Temperature Compensation 41°F to 95°F (5°C to 35°C)
Temperature Coefficient < ± ( 0.001%RD + 0.001%FS ) / °C outside of 5°C to 35°C

 

General Specifications
Specification Description
Channels Four total: one electrical, one internal pressure, two external pressure (measure only)
Enclosure IP Rating IP52 water and dust proof
Battery Rechargeable Li-Ion battery, typically 10 hours of operation, recharges in less than 4 hours
Display Color 800 x 480 TFT 5-inch touch screen
Communications USB, WiFi
Weight <4 lbs (<1.8 kg)
Size 9.3 x 4.3 x 2.8 in (235 x 110 x 70 mm)
Certification ISO17025 accredited certificate with data included
HART Communications Optional (ADT760-X-DL model)
Data Logging Optional (ADT760-X-DL model), up to 1,000,000 readings (date and time stamped)
Task Documentation Optional (ADT760-X-DL model) up to 1000 tasks
Automation Functions Switch test, auto step, leak test
Misuse Protection Up to 30 V on any two sockets
Multi Lingual Interface English, German, French, Italian, Spanish, Portuguese, Chinese, Japanese, and Russian
Pump Life >500,000 cycles
Power Rechargeable Li-Ion battery, external power: 110/220 V power adapter 10 V
Environment Specifications Operation: 32°F to 122°F (0°C to 50°C), 0-90% RH, less than 3,000 meters Compensated temperature: 32°F to 122°F (0°C to 50°C) Storage temperature: -4°F to 158°F (-20°C to 70°C)
Vibration and Shock Vibration: 4g (20 to 2,000 Hz) Shock: 8g, 1 meter drop test
Compliance CE
Software ACal, PCal, Land, LogII
Warranty 1 year

* See Internal Pressure Model Information in the Specifications Tab

Accessories (included)
Accessory Amount Picture
  1 set (to 1/4BSPF, to 1/4NPTF, to M20F)
USB to USB cable (ADT760-X-DL only) 1 pc
9816-X (110V/220V external power adapter) 1 pc
9724 (chargeable Li-ion battery) 1 pc
9025 (test leads for calibrator) 1 sets (3 pcs)
ADT100-760 Hose 1 pc for the ADT760-MA
Pressure Hose 1 pc for the ADT760-D
Silicone Tube 2 meters for the ADT761A-LLP, 1 meter for the ADT761A-D
1220211206 (Adapter, quick female to barb) (Only for ADT760-D) 1 pc
CD Manual 1 pc  
ISO17025 accredited certificate 1 pc  

 

Optional Accessories
Accessory Description Picture
ADT161 Intelligent Digital Pressure Module
9060 Pressure module connection cable
ADT100-760-N Special Connector to 1/4NPT quick connector
ADT100-760-B Special Connector to 1/4BSP quick connector
ADT100-760-M Special Connector to M20x1.5 quick connector
ADT100-760-N2 Special Connector to 1/2NPT quick connector
ADT100-760-B2 Special Connector to1/2BSP quick connector
ADT128-B Pneumatic Pressure Manifold for ADT760, -15 to 3000 psi, single port, 1/4 BSP F
ADT128-B2 Pneumatic Pressure Manifold for ADT760, -15 to 3000 psi, single port, 1/2 BSP F
ADT128-N Pneumatic Pressure Manifold for ADT760, -15 to 3000 psi, single port, 1/4 NPT F
ADT128-N2 Pneumatic Pressure Manifold for ADT760, -15 to 3000 psi, single port, 1/2 NPT F
ADT128-M Pneumatic Pressure Manifold for ADT760, -15 to 3000 psi, single port, M20x1.5
9240 DP gauge holder, rubber
1220211087 Filter, set of 1 pc
9913-760-SC Soft carrying case for ADT760, test leads, and many accessories
9914-760 Carry case for ADT760 and various accessories
ADT100-760-CNT Special connector (to be used to adapt from the ADT760 to ADT100-760-KIT adapters)
 

 