• Order with calibration
  • Not sure what you need? Ask our Team

Additel 780

Key Features

Enquire

For years, we’ve provided the most durable, accurate, quality pressure calibration products for field applications. The Additel 780 controller incorporates the same durability, accuracy, and quality into a new bench top controller packed with features and functionality that is remarkably easy to use. The Additel 780 offers two base ranges: to 1,000 psi (70 bar) and to 3,000 psi (200 bar). The base range establishes the maximum controlling range of the controller. Each configuration includes a control sensor in which you can preselect the sensor range best suited your application. External and internal sensors can be used which allows for expanded range and accuracy capability in the future.

Modular Design
Each unit comes with one Intelligent Pressure Module configured to the many range offerings provided. Standard accuracy sensors (ADT160A-CPXXX) are silicon pressure sensors with a 1 year accuracy of 0.02% FS. The precision quartz based sensors improve the 1 year accuracy specification to 0.005% of reading + 0.005% FS. Each sensor has been specially aged, tested and screened before assembly. After assembly each sensor is temperature compensated over the range of -10°C to 50°C. The Additel 780 allows for one internal pressure sensor and one external pressure sensor. The modular design of this unit provides for interchangeability of both the internal and external sensors with other Intelligent Pressure Modules.In addition to the Intelligent Pressure Modules, the Additel 780 has a built in barometric sensor. This allows for switching between gauge pressures to absolute pressures.

Standalone Solution
Typical pressure controllers will require a nitrogen bottle for the gas supply which make it difficult to move the controller around without having to move or connect to another bottle. The Additel 780 is unique in that with the optional electric pump, you can generate pressures to 1,000 psi (70 bar) without the need of a gas bottle.

Process Functionality
The 780 Pressure Controller is considered to handle a very wide range of applications which may normally require a pressure calibrator. Built-in capability, includes current and voltage measurement capability, 24 volt loop power, HART® and Profibus PA communication, switch measurement capability, selectable analog output and much more.

Specification ADT780-1K ADT780-3K
Pressure range -14.5 to 1,000 psi (-0.95 to 70 bar) -14.5 to 3,000 psi (-0.95 to 200 bar)[1]
Control stability 0.003% FS (stability based on % FS of control sensor range)
Precision (includes 1 year stability) See pressure range table
Media Clean gas
Over-range indication 103% to 120% (based on sensor)
Resolution 4, 5, 6, or 7 digits (user selectable)[2]
Pressure type Gauge, Absolute
Warm up time 15 minutes
Measurement units Pa, hPa, kPa, MPa, bar, mbar, psi, mmHg@0°C,cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C, mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C, kg/m2, kg/cm2, mtorr, torr, atm, lb/ft2, tsi, user selectable
Minimum control pressure[3] 0.0001 psi 0.001 psi

 

BAROMETRIC MEASUREMENT SPECIFICATION
Sensor/Mode Gauge Absolute
ADT160-02-CPXX N/A 40 Pa[1]
ADT160-01-APXXQ 3 Pa[2] N/A
Barometer range (60~110)kPa, the accuracy is 40 Pa (11 Pa is optional) [1] 40 Pa uncertainty (k=2) includes calibration uncertainty, linearity, and long term stability (<30 Pa per year). Barometer range of 60 to 120 kPa. [2] Combined linearity, hysteresis, and repeatability. Add 3 Pa when used in gauge mode. When using the ADT160-01-APXXQ sensors with the ADT780 controller in gauge mode, regular zeroing will realign the barometric reading to provide the most accurate result.

 

ELECTRICAL MEASURE SPECIFICATIONS
Specification Range Resolution Accuracy
Volts DC -30 to 30 V 0.1 mV ±0.01% rdg + 1.5 mV
-300 to 300 mV 1 μV ±0.01% rdg + 15 μV
Current DC -30 to 30 mA 0.1 μA ±0.01% rdg + 1.5 μA
Switch test If the switch has detected voltage, the range is from 3 – 24 V
DC 24V output 24V ± 0.24V, max 30 mA

 

PHYSICAL SPECIFICATIONS
Specification ADT780-1K & ADT780-3K
Power 100 to 240V, 50/60 Hz
Pressure ports G1/8 F
Storage temperature -20°C to 70°C
Operating environment 0-90% RH non-condensing
Display 7 inch (17.8 cm) color, touch screen display
Weight 33 lbs (15 kg)
Dimensions (DWH) 16.5 x 17.3 x 5.2 inch (419 x 440 x 132 mm)
Mounting Standard desktop, optional rack mount kit
Shock 4G
Vibration 1G 10Hz~500Hz

 

OTHER SPECIFICATIONS
WiFi specifications 802.11 b, g, and n
Vent Front panel vent and safety release button (only for 780S)
Communications RS232,USB,LAN,WIFI
Stability indicator User selectable
Operating modes Control, measure, and vent
Display modes Controller – show pressure indication and control
Localization English, Chinese (simplified), German, Spanish, French, Italian, Portuguese, Russian, Japanese
Conformity CE
Calibration certification Accredited certificate of calibration with data included
Warranty 1 year

 

Specifications for ADT780 Internal / External Sensors (Standard Pressure Sensors)
P/N Pressure Range Media Accuracy Pressure Rating Suggested Controller Compatibility Internal/ External Use
psig bar.g Burst Over Pressure
Standard Pressure Sensors
CP10 ±10 ±0.7 G 0.02% FS 3x 1.2x ADT780-1K only Both
CP15 ±15 ±1.0 G 0.02% FS 3x 1.2x ADT780-1K only Both
CP30 -15 to 30 -1 to 2.0 G 0.02% FS 3x 1.2x ADT780-1K only Both
CP50 -15 to 50 -1 to 3.5 G 0.02% FS 3x 1.2x ADT780-1K only Both
CP100 -15 to 100 -1 to 7.0 G,L 0.02% FS 3x 1.2x ADT780-1K only Both
CP300 -15 to 300 -1 to 20 G,L 0.02% FS 3x 1.2x Both Both
CP500 -15 to 500 -1 to 35 G,L 0.02% FS 3x 1.2x Both Both
CP600 -15 to 600 -1 to 40 G,L 0.02% FS 3x 1.2x Both Both
CP1K -15 to 1,000 -1 to 70 G,L 0.02% FS 3x 1.2x Both Both
CP2K -15 to 2,000 -1 to 140 G,L 0.02% FS 3x 1.2x ADT780-3K only Both
CP3K -15 to 3,000 -1 to 200 G,L 0.02% FS 3x 1.2x ADT780-3K only Both
CP5K -15 to 5,000 -1 to 350 G,L 0.02% FS 3x 1.2x N/A External
CP10K -15 to 10,000 -1 to 700 G,L 0.02% FS 2x 1.2x N/A External

 

Specifications for ADT780 Internal / External Sensors (Precision Sensors)
P/N
Media
Compound Pressure Absolute Pressure [3] Pressure Rating Suggested Controller Compatibility Internal/ External Use
Compound Pressure Range (psi, bar) Accuracy [2] Absolute Pressure Range (psi.a, bar.a) Accuracy Burst Over Pressure
Precision Sensors
CP30M G,L -15 to 30 psi (-1 to 2.0 bar) 0.01% rdg or 0.003%FS whichever is greater 0 to 45 psi.a (0 to 3 bar.a) 0.01% rdg or 0.003% FS whichever is greater[4], or 0.02%FS[5] 2x 1.2x ADT780-1K only Both
CP50M G,L -15 to 50 psi (-1 to 3.5 bar) 0.01% rdg or 0.003%FS whichever is greater 0 to 65 psi.a (0 to 4.5 bar.a) 0.01% rdg or 0.003% FS whichever is greater[4], or 0.02%FS[5] 2x 1.2x ADT780-1K only Both
CP100M G,L -15 to 100 psi (-1 to 7.0 bar) 0.01% rdg or 0.003%FS whichever is greater 0 to 115 psi.a (0 to 8 bar.a) 0.01% rdg or 0.003% FS whichever is greater[4], or 0.02%FS[5] 2x 1.2x ADT780-1K only Both
CP300M G,L -15 to 300 psi (-1 to 20 bar) 0.01% rdg or 0.003%FS whichever is greater 0 to 315 psi.a (0 to 21 bar.a) 0.01% rdg or 0.003% FS whichever is greater 2x 1.2x Both Both
CP500M G,L -15 to 500 psi (-1 to 35 bar) 0.01% rdg or 0.003%FS whichever is greater 0 to 515 psi.a (0 to 36 bar.a) 0.01% rdg or 0.003% FS whichever is greater 2x 1.2x Both Both
CP1KM G,L -15 to 1000 psi (-1 to 70 bar) 0.01% rdg or 0.003%FS whichever is greater 0 to 1015 psi.a (0 to 71 bar.a) 0.01% rdg or 0.003% FS whichever is greater 2x 1.2x Both Both
CP1.5KM G,L -15 to 1500 psi (-1 to 100 bar) 0.01% rdg or 0.003%FS whichever is greater 0 to 1515 psi.a (0 to 101 bar.a) 0.01% rdg or 0.003% FS whichever is greater 2x 1.2x Both Both
CP2KM G,L -15 to 2000 psi (-1 to 140 bar) 0.01% rdg or 0.003%FS whichever is greater 0 to 2015 psi.a (0 to 141 bar.a) 0.01% rdg or 0.003% FS whichever is greater 2x 1.1x  ADT780-3K only Both
CP3KM G,L -15 to 3000 psi (-1 to 200 bar) 0.01% rdg or 0.003%FS whichever is greater 0 to 3015 psi.a (0 to 201 bar.a) 0.01% rdg or 0.003% FS whichever is greater 2x 1.1x  ADT780-3K only Both

 

ADT780-1K-EPUMP
Pressure range -13.8 to 1,080 psi (-0.95 to 75 bar)
Weight 69.6 lbs (31.6 kg)
Size(WHD) 17.3 X 9.3 X 21.7 inch (440 X 235X 550 mm)
Power 100 to 240V
Media Air
Volume 150 mL
Outlet port G1/8F
Storage temperature -20°C to 70°C
Operating environment 0-90% RH non-condensing
Mounting Standard desktop, optional rack mount
Control operation 2 LED displays with pressure limit settings
Typical max pressure time Approx 2 min with ADT780 Controller

Sales Enquiry