• Order with calibration
  • Not sure what you need? Ask our Team

AccuMac AM1710

Key Features

Enquire

The AM1710 series of Secondary Reference PRTs offers affordable reference probes for precision temperature measurement and calibration both in the lab and in the field. These PRTs feature an accuracy of <0.012°C and a short term stability of <0.007°C.

To reach the best performance in stability and repeatability, the sensing element has been specially designed to protect the platinum sensing wire from contamination at high temperatures. A unique supporting structure and filling material provide the best balance among the hysteresis effect, mechanical shock and thermal shock performance. This high performance probe fully meets the ITS-90 criteria for reference thermometers. Each probe comes standard with an NIST-traceable certificate of calibration.

General Specifications
  AM1710-12 AM1710-BEND
Temperature Range -40°C to 160°C
Nominal Resistance at 0.01°C 100 Ω
Temperature Coefficient 0.003925 Ω/Ω/°C
Calibrated Accuracy (k=2) [1][2]
±0.025°C at -40°C
±0.015°C at 0.01°C
±0.025°C at 160°C
Long Term Drift at 0.01°C [3] <0.01°C at TPW after 100 hours at 160°C
Short Term Stability <0.007°C
Thermal Shock <0.005°C after 10 thermal cycles from minimum to maximum temperatures
Hysteresis ≤0.005°C
Self-heating 50 mW/°C
Response Time [4] 9 seconds
Sheath Length 12 in (305 mm) 7 in (178 mm) from tip to bend, 4 in (101.6 mm) from bend to handle
Sheath Diameter 0.25 in (6.35 mm)
Sensor ElementLength 1.26 in (32 mm)
Sensor Element Location 0.12 in (3 mm) from tip
Insulation Resistance >1000 MΩ at room temperature
Sheath Material Stainless steel
External Leads Teflon™ insulated copper wire, 4 leads, 6.5 feet (2 meters)
Handle Dimensions 0.59 in (15 mm) (OD) 0.59 in (15 mm) (OD)
X 2.56 in (65 mm) (L) X 2.56 in (65 mm) (L)
Handle Temperature Range 40°C to 160°C
Calibration NIST-Traceable calibration with data [5]
[1] Includes calibration and 100 hour drift.
[2] Probe calibration ranges may differ from probe temperature ranges (see Calibrated Accuracy for calibration ranges).
[3] Long-term drift rate is for reference only. It could be affected by such facts as handling, application, and maintenance, etc.
[4] For 63% response to step change in water moving at 1 meter per second.
[5] For 17025 accredited calibration, please contact Additel.

 

Optional Accessories
Model Quantity Description Picture of Optional Acccesory
9070 1 pc Smart connector for reference PRT for use with Additel temeprature products
 
9071 1 pc Connector adapter from Additel smart connector to 4-wire with gold plated spades
 
9072 1 pc Connector adapter from Additel smart connector to 4-wire with clamps
 

Sales Enquiry