• Order with calibration
  • Not sure what you need? Ask our Team

AccuMac AM1760/1762

Key Features

Enquire

The AM1760 series Secondary SPRTs provides our customers with reliable secondary standards that can be used daily in their labs. These SPRTs feature an accuracy of <0.006°C at 0.01°C, a short term stability of <0.002°C and a very low drift rate of less than 0.004°C after 100 hours at 661°C. Two different lengths of these SPRTs are available at either12 inches (305 mm) or 20 inches (508 mm).

The sensing element is designed to protect the platinum sensing wire from contamination at high temperatures, giving these SPRTs a high level of stability and repeatability in performance. A uniquely designed support structure and filling material provides excellent balance between the hysteresis effect, mechanical shock and thermal shock performance.

This high performance probe fully meets the ITS-90 criteria for reference thermometers. Each probe comes standard with an NIST-traceable certificate of calibration.

General Specifications  
  AM1760 AM1762 AM1760-BEND AM1762-BEND
Temperature Range -200°C to 670°C
Nominal Resistance at 0°C 100 Ω 25 Ω 100 Ω 25 Ω
Temperature Coefficient 0.003925 Ω/Ω/°C
Calibrated Accuracy (k=2) [1][2]
±0.010°C at -196°C
±0.006°C at 0.01°C
±0.015°C at 420°C
±0.025°C at 661°C
Long Term Drift at 0.01°C [3] <0.004°C at TPW after 100 hours at 661°C
Short Term Stability <0.002°C
Thermal Shock <0.002°C after 10 thermal cycles from minimum to maximum temperatures
Self-heating 0.0015°C at 0.5 mA current 0.0015°C at 1 mA 0.0015°C at 0.5 mA 0.0015°C at 1 mA
Response Time [4] 9 seconds
Measurement Current 0.5 mA 1 mA 0.5 mA 1 mA
Sensor Length 1.65 in (42 mm)
Sensor Location 0.2 in (5 mm) from tip
Insulation Resistance >1000 MΩ at room temperature
Sheath Material Inconel™
Sheath Dimensions 1760-20: 0.25 in (6.35 mm) (OD) x 20 in (508 mm) (L) 1762-20: 0.25 in (6.35 mm) (OD) x 20 in (508 mm) (L) 0.25 in (6.35 mm) (OD) x 9.75 in (247.65 mm) vertical x 4.5 in (114.3 mm) horizontal
1760-12: 0.25 in (6.35 mm) (OD) x 12 in (305 mm) (L) 1762-12: 0.25 in (6.35 mm) (OD) x 12 in (305 mm) (L)
External Leads Teflon™ insulated copper wire, 4 leads, 6.5 feet (2 meters)  
Handle Dimensions 0.59 in (15 mm) (OD) x 2.56 in (65 mm) (L)
Handle Temperature Range -50°C to 180°C  
Calibration NIST-Traceable calibration with data [5]  
[1]  Includes calibration and 100 hour drift.  
[2]  Probe calibration ranges may differ from probe temperature ranges (see Calibrated Accuracy for calibration ranges).  
[3]  Long-term drift rate is for reference only. It could be affected by such facts as handling, application, and maintenance, etc.  
[4]  For 63% response to step change in water moving at 1 meter per second.  
[5]  For 17025 accredited calibration, please contact Additel.  

Optional Accessories
Model Quantity Description Picture of Optional Acccesory
  1 pc Smart connector for reference PRT for use with Additel temeprature products
 
9071 1 pc Connector adapter from Additel smart connector to 4-wire with gold plated spades
 
9072 1 pc Connector adapter from Additel smart connector to 4-wire with clamps
 

Sales Enquiry

AccuMac AM1760/1762